gc9720plus

The most advanced GC from FULI.

Category:

Gas chromatograph gc9720plus

1. System capability

  • Assembly simultaneously: 2 inlets+3 detectors(FID, TCD, ECD, FPD  optional)
  • Independent split cold-trap and septum collecting trap,
    Carrier gas control mode: pressure, flow, programming pressure, programming flow,
    Programming pressure/flow/linear speed: max 8 steps,
    Maintenance tracking reminder, intelligent regular maintenance plan,

2. EPC/AFC gas system

  • Full AFC/EPC, 6 AFC module with 18-channel AFC/EPC independent control,
    AFC/EPC pressure accuracy 0.001psi, Retention time RSD <=0.02%, Peak area RSD <=1.0%,
  • Carrier and makeup gas settings selectable for He, H2, N2 and Ar
  • Psi, KPa, Bar units selectable
  • Pressure control range: 1~600KPa, Pressure accuracy: 0.01Psi, Programmable pressure ramp RSD ≤0.5%
  • Flow rate control range: 0~600ml/min, flow rate accuracy: 0.1ml/min, flow rate RSD≤0.1%
  • Flow sensor accuracy: ≤±3%, detector module accuracy: ≤±7%

 

 

 

 

 

 

 

 

 

 

 

 

 

3. Column oven—Excellent temperature control system

  • Column oven dimension: 278×310×165mm=14.2L, accommodate up to 2pcs 105m×0.53mm ID capillary column
  • Temp. control range: RT+5℃~450℃(0.1℃ increment)
  • Standard cool-down mode 450—50℃<=5min
  • Optional quick cool-down mode 450—50℃<=3.5min
  • Scientific inner heating interchange to achieve high consistency in between front and back column heating distribution
  • Reproducible temp. ramp process in column oven to achieve good retention time RSD
  • Max temp.-ramp rate: 120 ℃/min
  • Programming temp.-ramp: 30 steps

GC9720 oven

 

 

 

 

 

 

 

 

 

 

 

4. Inlets

  • Packed purge injection port(PPIP)
  • Split/splitless capillary port(S/SL)
    1. Max temperature: 450℃
    2. Split ratio: 4500:1
    3. Gas saver mode to reduce gas consumption without compromising performance
    4. Programmable temp./pressure ramp: 8 steps
    5. Manual dismount in 10s to make liner replacement easier
    6. Intelligent tip to record and remind replacement/maintenance of septum, liner and split cold-trap

 

 

 

 

 

5. Detectors

DetectorMax operating temp.
Limit of detection
Baseline noise
Baseline drift
(after 2hrs stabilization)
Linear dynamic range
FID
450℃
≤3.0×10^-12g/s
(N-C16)
≤2×10^-14 A
5×10^-14 A/30 min
≥10^7
TCD
400℃≥10000mV.ml/mg
(N-C16)
≤30uV
≤100uV/30min
≥10^4
ECD
400℃≤ 3×10^-14g/ml
(γ-666)
≤20uV
≤50uV/30min
≥10^4
FPD
400℃S: ≤2.0×10^-11g/s
P: ≤5.0×10^-13g/s
or 2.0×10^-13g/s
S: ≤2×10^-13A
P: ≤8×10^-13A
S: ≤1×10^-12 A/30min
P: ≤2×10^-12 A/30min
S: ≥10^2
P: ≥10^3

6. Autosampler

  • Single FL1092T, 16-vial autosampler
  • Double FL1092T, double-tower independent asynchronous control or synchronous control to improve analysis efficiency
  • Single/double FL1092T+FL1092C, 150-vial autosampler, heating/barcode scan/stirring and whole plate cooling/heating are optional

 

 

 

 

 

 

 

 

7. Multi-valve system

  • Multi-valve/multi-column available to customize per different application, max 4 external valve can be installed
  • One-time injection for complete gas analysis
  • High stability and low valve switchover fluctuation

 

 

 

 

 

 

 

8. Workstation and data communications

  • LAN
  • Windows 10/7 support
  • Signal capture from 2 detectors simultaneously
  • Sampling frequency: 100Hz
  • Sampling speed: 50 times/sec
  • Unlimited peak quantity
  • Self-diagnosis: intelligent automatic error identification and self-protection
  • FL97PLUS software and Clarity V8.3 to fully support FDA-21 CFR PART 11

 

 

 

 

 

 

 

 

 

 

 

 

9. Basic info

  • Votage: 220V±10%, 50Hz
  • Power: 3000W
  • Dimension:  645×500×555mm
  • Net weight: 55KG